15-Day
Refurbishment cadence
OEM-Spec
Verified against OEM and published benchmarks; report included.
On-Staff
Solution Scientists
Since 1996
Three decades supporting labs
What you are buying
Environmental, forensic, and food-safety methods that identify and confirm compounds by mass run on the GC-MS the Agilent 5977A MSD delivers. This system pairs the 7890 GC and 5977A MSD with both a 7697 headspace sampler and a 7693 liquid autosampler, sharing one GC inlet through a transfer line interface accessory. The headspace sampler pulls vapor from sealed vials for volatile analysis; the liquid autosampler injects directly for routine compounds. GenTech rebuilds and tests the GC, the samplers, the source, and the analyzer, then verifies the system end to end.
Every unit moves through the GenTech Certified process. Solution Scientists disassemble and inspect each module, replace all consumables and degraded wear components, then tune and calibrate the instrument and verify performance against OEM and published benchmarks using known reference chemicals. The instrument ships with its verification report. A certified-refurbished system gives you that performance without the capital cost or the lead time of a new purchase.
What Agilent says about this platform
The 7697A Headspace Sampler builds on the 7890 GC and 7693A automatic liquid sampler platform, available in a 12-vial configuration for lighter sample loads or a 111-vial configuration with three priority positions for late-arriving samples. A transfer line interface accessory lets the headspace sampler and the ALS share the same GC inlet. An automatic vial leak test checks each vial during pressurization without a separate manual calibration step. Independent control of vial pressurization and column head pressure keeps back pressure from affecting sample transfer. Multiple headspace concentration mode runs up to 100 extractions per GC start; multiple headspace extraction mode runs up to 100 extractions per vial. Temperature zones for the oven, the valve and loop, and the transfer line each set in 1-degree increments with 0.1-degree resolution. The sampler supports 10, 20, or 22-milliliter vials without an adapter and runs under OpenLab CDS, MassHunter, or Waters Empower software.
What each component does
Gas chromatograph
Separates the mixture on a capillary column under programmed temperature and electronic pneumatic control. The 7890 front end runs the same inlets, detectors, and Capillary Flow Technology as a standalone GC.
Headspace sampler
The 7697 pulls vapor from a sealed vial for volatile analysis, sharing the GC inlet with the liquid autosampler through a transfer line interface accessory.
Liquid autosampler
The 7693 draws a precise volume from a vial and injects it directly onto the column for routine, non-volatile compounds.
Transfer line and ion source
Carries the separated compounds into the 5977A MSD at controlled temperature. Electron ionization produces reproducible, library-searchable spectra; chemical ionization preserves the molecular ion for confirmation.
Mass analyzer
The quadrupole filters ions by mass. It scans a full spectrum for identification or holds on selected ions for the lowest detection limits in quantitation.
| In this system | |
| Agilent 7890B G3440A | Verified |
| 5977A MSD G4514A | Verified |
| 7683B Autosampler G2614A | Verified |
| Optional Inlets Optional Detectors | Verified |
|
Workstation loaded with software
|
Included |
| Each component is individually serialized and performance-verified before the system is assembled and tested as a complete system. | |
Technical detail
| 7697 | Headspace Sampler |
| Vial capacity | 12-vial or 111-vial (3 priority) config |
| Extraction modes | MHC / MHE, up to 100 extractions |
| Temperature zones | Oven, valve/loop, transfer line, 1°C steps |
| G4514A | 5977A MSD |
| Mass range | 1.6 to 1,050 u |
| Software | Loaded on workstation |
*OEM specifications for brand new instrument.
What “GenTech Certified” means
Refurbishment at GenTech is a defined process, not a wipe-down. Every system moves through the same five stages.
1. Disassembly and rebuild. Solution Scientists strip the instrument down, clean each module inside and out, and rebuild it from the frame up. They replace pump seals, pistons, and check valves to specification.
2. Configuration and new parts. Solution Scientists install the options and modules you selected and fit new wear parts throughout. They replace the detector lamp with a new one and reset the lamp hours.
3. Diagnostic and performance testing. Solution Scientists tune and calibrate the instrument, run system diagnostics, and confirm detector wavelength accuracy. They test full-system pressure and gradient performance against OEM and published benchmarks using known reference chemicals.
4. Data package you approve before shipment. The verification data goes to you for approval before the system leaves the bench. You see exactly how the instrument performs before it ships.
5. Delivery and installation. The instrument ships in protective packing with a one-year warranty option and professional installation available from a GenTech Solution Scientist.
Labs that run this system
The Agilent 7890 GC with 5977A MSD, 7697 Headspace Sampler & 7693 AS is method-compatible across the techniques most regulated and research labs run every day. These are the application areas buyers ask about most.
Hydrocarbons and fuels
Solvents and fatty acids
Permanent and fixed gases
Pesticides and halogenated compounds
Nitrogen and phosphorus pesticides
Sulfur and phosphorus compounds
Volatiles and semivolatiles
Residual solvents confirmation
Pesticides and terpenes
| Weight | 350 lbs |
|---|---|
| Dimensions | 60 × 60 × 60 in |
| Condition | Refurbished |
| Model | 5977, 7890, 7697, 7693 |
| OEM | Agilent Technologies |
| Financing | Financing Options Available |
Put the financing content and link here.
Need a single detector instead?
Chimney won’t light or expanding an existing system? Every part in this system is available individually, in stock and performance-verified.







