What you are buying
A refurbished Agilent 1260 I VWD, a variable wavelength detector. It measures absorbance at selected wavelengths with high sensitivity.
Every unit moves through the GenTech Certified process. Solution Scientists disassemble and inspect each module, replace all consumables and degraded wear components, then tune and calibrate the instrument and verify performance against OEM and published benchmarks using known reference chemicals. The instrument ships with its verification report. A certified-refurbished system gives you that performance without the capital cost or the lead time of a new purchase.
What Agilent says about this platform
Agilent built the G1314 family as a double-beam photometer with a deuterium lamp, covering 190 to 600 nanometers. Short-term noise runs about 0.75 x 10-5 AU at 254 nanometers. Drift holds to roughly 3 x 10-4 AU per hour at the same wavelength. Linearity extends to 2 AU before the response departs from proportional. A built-in holmium oxide filter verifies wavelength accuracy without an external standard. Programmable wavelength switching and stop-flow scanning let an operator retune detection to each analyte’s elution or scan a spectrum without restarting the run. The standard flow cell holds 14 microliters at a 10-millimeter path length and up to 40 bar. Early Maintenance Feedback tracks lamp burn time against a user-set limit and keeps an electronic record of maintenance and errors.
What each module does
Variable wavelength detection
Measures absorbance at selected wavelengths with high sensitivity, over a range of roughly 190 to 600 nm. It drops into a compatible 1260 Infinity HPLC stack under OpenLab CDS or ChemStation.
| In this stack | |
| Agilent 1260 VWD G1314B / C / D / E / F | Verified |
| 1260 Vialsampler G7129A | Verified |
| 1260 Multicolumn Thermostat G7116A | Verified |
| 1260 Diode Array Detector G7115A | Verified |
|
Workstation loaded with software
|
Included |
| Each module is individually serialized and performance-verified before the stack is assembled and tested as a complete system. | |
Technical detail
| Detection type | Double-beam photometer, deuterium lamp |
| Wavelength range | 190 to 600 nm |
| Short-term noise | ~0.75 x 10-5 AU at 254 nm |
| Drift | ~3 x 10-4 AU/hr at 254 nm |
| Flow cell | 14 µL, 10 mm path, 40 bar max |
| Software | Loaded on workstation |
*OEM specifications for brand new instrument.
What “GenTech Certified” means
Refurbishment at GenTech is a defined process, not a wipe-down. Every system moves through the same five stages.
1. Disassembly and rebuild. Solution Scientists strip the instrument down, clean each module inside and out, and rebuild it from the frame up. They replace pump seals, pistons, and check valves to specification.
2. Configuration and new parts. Solution Scientists install the options and modules you selected and fit new wear parts throughout. They replace the detector lamp with a new one and reset the lamp hours.
3. Diagnostic and performance testing. Solution Scientists tune and calibrate the instrument, run system diagnostics, and confirm detector wavelength accuracy. They test full-system pressure and gradient performance against OEM and published benchmarks using known reference chemicals.
4. Data package you approve before shipment. The verification data goes to you for approval before the system leaves the bench. You see exactly how the instrument performs before it ships.
5. Delivery and installation. The instrument ships in protective packing with a one-year warranty option and professional installation available from a GenTech Solution Scientist.
Labs that run this system
The 1260 HPLC is method-compatible across the techniques most regulated and research labs run every day. These are the application areas buyers ask about most.
Assay, Impurity and dissolution
The 1260 transfers cleanly into USP monograph assays, related-substance profiling, and dissolution testing. Quaternary blending simplifies mobile-phase transfer between labs.
EPA HPLC methods
Diode array detection supports EPA HPLC methods for contaminants and residues in water and soil, including PAHs, explosives, and carbamates.
Additives and vitamins
The system quantifies preservatives, sweeteners, organic acids, and fat-soluble vitamins at the throughput a production-support lab needs.
Cannabinoid potency
HPLC with UV detection is the reference technique for potency. The 1260 resolves the major cannabinoids for compliance and label-claim testing.
Botanical actives
Labs quantify polyphenols, flavonoids, and marker compounds in extracts and nutraceutical formulations across long, reproducible sequences.
Screening and drug monitoring
The 1260 runs HPLC-UV workflows for therapeutic drug monitoring and drugs-of-abuse screening, where method portability and uptime matter more than peak novelty.
Purity and additives
Industrial labs separate and quantify synthesis intermediates, polymer additives, and stabilizers for QC and release testing.
Four-solvent scouting
Agilent recommends the quaternary pump for method development. Blend up to four solvents to scout selectivity before you lock a final method.
Proteomics, metabolomics, lipidomics
Many omics workflows use mass spectrometry for detection and rely on LC for separation. The 1200 serves as the front end feeding your MS or fraction collector.
| Condition | Refurbished |
|---|---|
| Model | G1314F |
| OEM | Agilent Technologies |
| Financing | Financing Options Available |
Financing Options Available. Finance Builder.
Need a single module instead?
Replacing a failed pump or expanind an existing stack? Every module in this system is available individually, in stock and performance-verified.